Skip to content

Public FDA record

Z-1975-2021: PERLA TL Lateral Connector Open - Product Usage: intended to be used with autograft and/or allograft. Ped

Recall record Z-1975-2021 for SPINEART SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1975-2021
Company in FDA record
SPINEART SA
FDA record creation date
Jun 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to external labelling error/mix-up. Affected product external labels list the incorrect rod size.

Source facts

Action
On 04/12/2021, the firm sent an email to its US Distribution Center and a follow-up "URGENT: MEDICAL DEVICE RECALL" letter (dated April 14, 2021) TO inform customers that the recall was being initiated due to an external labelling error where the external labels were mixed-up for two batches and could lead to confusion for the hospital staff and induce a minimal surgery delay. Customers are instructed to: 1. Immediately review their inventory and quarantine concerned products if any. 2. If they may have further distributed this product; please identify concerned customers and notify them at once of this product recall by using this document. 3. Collect and quarantine all products. 4. Sent back all products with the enclosed Response Form to Spineart Geneva (Chemin du Pr Fleuri 3, 1228 Plan-les-Ouates, Switzerland). E-mail: regulatory@spineart.com. 5. All returned products will be exchanged with batches already available in our warehouse.
Initiated
2021-04-09
Root cause
Process control
Terminated
2022-09-29
Distribution
U.S. Nationwide distribution in the states of CA, MA, and NC. O.U.S.: Not provided
Recall number
Z-1975-2021
Recall status
Terminated
CDRH record ID
187567
FDA record created
2021-06-28
Recall event number
87971

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗