Z-1975-2023: Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30

Recall record Z-1975-2023 for Helena Laboratories, Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1975-2023
Company in FDA record
Helena Laboratories, Corp.
FDA classification date
Jun 16, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Packaging for a serrated blade applicator kit (12 Sample) may contain a serrated blade applicator (18 sample). This results in a characteristic pattern of missed or light samples on the gel, leading to incorrect or no result, or delay in result.

Source facts

Action
On April 27, 2023, Helena Laboratories contacted consignees via phone to make them aware of the recall. On May 2, 2023, Helena Laboratories followed up by issuing an Urgent Medical device Recall notification via E-Mail. Helena Laboratories asks consignees to take the following actions: 1. Carefully open each box while wearing gloves. Do not touch the teeth of the applicator as this can affect its use during sample delivery. 2. Locate the number imprinted on the handle of the applicator. The number may face outward or inward depending on the placement in the box. Incorrect applicators are imprinted with the number 18. 3. Remove and discard any boxes with applicators imprinted with the number 18. If the applicator is imprinted with the number 12, mark the box to indicate that it has been examined and contains the correct product. 4. Notate the total number of boxes inspected on the attached form. Affected products will be credited or replaced. 5. If you have further distributed this product, please identify your customers and notify them at once of this recall. Your notification to your customers may include a copy of this recall notification letter. This recall should be carried out at the user level. 6. Please complete and return the enclosed Acknowledgement and Receipt Form to technicalservice@helena.com or mail.
Initiated
2023-05-02
Root cause
Process design
Distribution
Distribution to US states of PA, NE, MI, FL, AR, NY, TX, and CA, and to Thailand.
Recall number
Z-1975-2023
Recall status
Open, Classified
CDRH record ID
200156
FDA classified
2023-06-16
Recall event number
92297

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗