There is a potential for an assembly error resulting in the battery wire being connected to the ON/OFF power switch incorrectly.
Source facts
Action
On June 24, the firm plans to send URGENT MEDICAL DEVICE letters to six consignees that will advise them of the issue, request that they either return their device or perform an ON/OFF switch verification procedure, and to return the attached REPLY FORM.
Initiated
2019-06-24
Root cause
Process control
Terminated
2021-04-01
Distribution
United States - AR, NM, PA, TX, and WA.
Recall number
Z-1976-2019
Recall status
Terminated
CDRH record ID
173926
FDA record created
2019-07-15
Recall event number
83173