Open original FDA record ↗Recall record
- FDA identifier
- Z-1977-2014
- Company in FDA record
- BD Biosciences, Systems & Reagents
- FDA classification date
- Jul 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Laser cooling fans and internal fan of the associated fluidics cart are not operational.
Source facts
- Action
- BD sent an Important Product Recall notification letters dated June 2014 to affected US customers by trackable mail. International affiliates were notified on June 6, 2014. The letter identified the affected product, problem and the actions to be taken. A Field Change notification has been established and a BD field Service Engineer will coordinate with each account to perform the necessary corrections. Questions or concerns may be directed to BD Customer Support at BD Biosciences, prompt 2. Out of US customers are to contact local BD rep with questions.
- Initiated
- 2014-06-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-01-27
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, OH, RI, GA, FL, MD, NY, WV, NJ, Ma, CT, IA, NM, MN and the countries of China and Belgium.
- Recall number
- Z-1977-2014
- Recall status
- Terminated
- CDRH record ID
- 128033
- FDA classified
- 2014-07-02
- Recall event number
- 68490