Skip to content

Public FDA record

Z-1977-2014: BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnost

Recall record Z-1977-2014 for BD Biosciences, Systems & Reagents. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1977-2014
Company in FDA record
BD Biosciences, Systems & Reagents
FDA classification date
Jul 2, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Laser cooling fans and internal fan of the associated fluidics cart are not operational.

Source facts

Action
BD sent an Important Product Recall notification letters dated June 2014 to affected US customers by trackable mail. International affiliates were notified on June 6, 2014. The letter identified the affected product, problem and the actions to be taken. A Field Change notification has been established and a BD field Service Engineer will coordinate with each account to perform the necessary corrections. Questions or concerns may be directed to BD Customer Support at BD Biosciences, prompt 2. Out of US customers are to contact local BD rep with questions.
Initiated
2014-06-06
Root cause
Nonconforming Material/Component
Terminated
2015-01-27
Distribution
Worldwide Distribution - US Nationwide in the states of CA, OH, RI, GA, FL, MD, NY, WV, NJ, Ma, CT, IA, NM, MN and the countries of China and Belgium.
Recall number
Z-1977-2014
Recall status
Terminated
CDRH record ID
128033
FDA classified
2014-07-02
Recall event number
68490

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗