Skip to content

Public FDA record

Z-1978-2021: NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model

Recall record Z-1978-2021 for NeuMoDx Molecular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1978-2021
Company in FDA record
NeuMoDx Molecular Inc
FDA record creation date
Jun 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for false positive results when certain lots of cartridges are used in conjunction with specific assays.

Source facts

Action
URGENT MEDICAL DEVICE CORRECTION notification letters dated 5/4/21 were sent to customers. Actions Required by Customers 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Medical Device Correction Notice with all users of the NeuMoDxTM Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDxTM Cartridges from any of these lots in combination with the NeuMoDxTM SARS-CoV-2 Assay or the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay, please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this medical device correction. 4. Complete the attached response form and return to Quality.Communications@qiagen.com to confirm receipt of this notification. We sincerely apologize for any inconvenience this may have caused and thank you in advance for your patience and cooperation. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com
Initiated
2021-05-04
Root cause
Material/Component Contamination
Terminated
2022-06-21
Distribution
Worldwide distribution.
Recall number
Z-1978-2021
Recall status
Terminated
CDRH record ID
187860
FDA record created
2021-06-29
Recall event number
88099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.