Open original FDA record ↗Recall record
- FDA identifier
- Z-1978-2021
- Company in FDA record
- NeuMoDx Molecular Inc
- FDA record creation date
- Jun 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for false positive results when certain lots of cartridges are used in conjunction with specific assays.
Source facts
- Action
- URGENT MEDICAL DEVICE CORRECTION notification letters dated 5/4/21 were sent to customers. Actions Required by Customers 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Medical Device Correction Notice with all users of the NeuMoDxTM Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDxTM Cartridges from any of these lots in combination with the NeuMoDxTM SARS-CoV-2 Assay or the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay, please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this medical device correction. 4. Complete the attached response form and return to Quality.Communications@qiagen.com to confirm receipt of this notification. We sincerely apologize for any inconvenience this may have caused and thank you in advance for your patience and cooperation. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com
- Initiated
- 2021-05-04
- Root cause
- Material/Component Contamination
- Terminated
- 2022-06-21
- Distribution
- Worldwide distribution.
- Recall number
- Z-1978-2021
- Recall status
- Terminated
- CDRH record ID
- 187860
- FDA record created
- 2021-06-29
- Recall event number
- 88099