Skip to content

Public FDA record

Z-1978-2026: Halyard SAMMC ANGIOGRAPHY kit. Model Numbers: SAMM066-15.

Recall record Z-1978-2026 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1978-2026
Company in FDA record
AVID Medical, Inc.
FDA classification date
May 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Source facts

Action
Owens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure. All usage guidelines and the Instructions for Use are provided in Medline s recall letter. Consignees were also instructed to complete and return the Recall Response Form as well as to notify customers if affected kits were further distributed to ensure all end-users are appropriately notified and maintain records of effectiveness.
Initiated
2026-03-13
Root cause
Under Investigation by firm
Distribution
US distribution to FL, IL, MO, MS, NC, NE, TX.
Recall number
Z-1978-2026
Recall status
Open, Classified
CDRH record ID
219447
FDA classified
2026-05-01
Recall event number
98686

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗