Open original FDA record ↗Recall record
- FDA identifier
- Z-1979-2013
- Company in FDA record
- Boston Scientific CRM Corp
- FDA classification date
- Aug 16, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Boston Scientific CRM manufacturing quality system recently discovered test artifacts (related to shock charge time) in a small number of implantable defibrillators (ENERGEN DR ICD model E143) which require further investigation. While these devices successfully passed all manufacturing tests and met all requirements, they showed a slightly longer charge time than other devices during a manufactu
Source facts
- Action
- Consignee was sent on 7/15/13 a Boston Scientific "Medical Device Retrieval" letter dated July 12. 2013. The letter described the product problem and informed the consignee that they will replace the product removed from their inventory. For further questions, contact Boston Scientific Technical Services at 1 800 227-3422.
- Initiated
- 2013-07-15
- Root cause
- Under Investigation by firm
- Terminated
- 2013-11-12
- Distribution
- US distribution in the state of Tennessee.
- Recall number
- Z-1979-2013
- Recall status
- Terminated
- CDRH record ID
- 120194
- FDA classified
- 2013-08-16
- Recall event number
- 65798