- FDA identifier
- Z-1979-2014
- Company in FDA record
- Nihon Kohden America Inc
- FDA classification date
- Jul 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Source facts
- Action
- Nihon Kohden sent an Urgent Medical Device Safety Notification and Field Correction letter dated June 3, 2014 to affected customers. Recall notification letter, dated June 3, 2014 was hand-delivered to site by Nihon personnel. The letter provided the following information: product description with codes and software version, reason for recall, field correction information and actions to be taken, MedWatch reporting information, and contact information for recall.
- Initiated
- 2014-06-03
- Root cause
- Software design
- Terminated
- 2015-04-28
- Distribution
- Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
- Recall number
- Z-1979-2014
- Recall status
- Terminated
- CDRH record ID
- 127799
- FDA classified
- 2014-07-02
- Recall event number
- 68459