Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.
Source facts
Action
Urgent Medical Device Correction notification letters dated 6/3/19 were sent to customers.
Initiated
2019-06-03
Root cause
Material/Component Contamination
Terminated
2020-08-28
Distribution
The products were distributed to the following US states: AL, IN, and KY.
Recall number
Z-1979-2019
Recall status
Terminated
CDRH record ID
173499
FDA record created
2019-07-16
Recall event number
83095