Reason stated in the source
The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.
Source facts
- Action
- The firm, Elekta, disseminated their "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" by email on 06/24/2019 to its customers. The notices identified the problem, provided recommended actions, and requested the notice be placed in a place accessible to all users. Customers were also instructed to complete and return the Acknowledgement Form. A software update will be provided once available. If you have any questions, contact the Post Market Surveillance Specialist by email: Elekta_Global_PMS@elekta.com.
- Initiated
- 2019-06-24
- Root cause
- Software design
- Terminated
- 2021-04-01
- Distribution
- US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.
- Recall number
- Z-1980-2019
- Recall status
- Terminated
- CDRH record ID
- 173856
- FDA classified
- 2019-07-16
- Recall event number
- 83158