Skip to content

Public FDA record

Z-1980-2025: GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Recall record Z-1980-2025 for Ohio Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1980-2025
Company in FDA record
Ohio Medical Corporation
FDA classification date
Jun 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

Source facts

Action
Ohio Medical issued a Field Safety Notice to its consignees on 05/08/2025 followed by telephone calls on 05/13/2025. The notice explained the problem with the device, potential risk, and requested the devices be returned for correction. The notice further stated: "The Field Safety Notice is communicated to all impacted customers and is communicated to competent authorities in the affected markets. Receivers should pass this FSN on to all persons with a need to be advised from within your organization or to any organization where the potentially affected products have been supplied. Please consider end users and others within the supply chain in the circulation of this notice."
Initiated
2025-05-08
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.
Recall number
Z-1980-2025
Recall status
Open, Classified
CDRH record ID
213781
FDA classified
2025-06-13
Recall event number
96851

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗