Open original FDA record ↗Recall record
- FDA identifier
- Z-1980-2025
- Company in FDA record
- Ohio Medical Corporation
- FDA classification date
- Jun 13, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
Source facts
- Action
- Ohio Medical issued a Field Safety Notice to its consignees on 05/08/2025 followed by telephone calls on 05/13/2025. The notice explained the problem with the device, potential risk, and requested the devices be returned for correction. The notice further stated: "The Field Safety Notice is communicated to all impacted customers and is communicated to competent authorities in the affected markets. Receivers should pass this FSN on to all persons with a need to be advised from within your organization or to any organization where the potentially affected products have been supplied. Please consider end users and others within the supply chain in the circulation of this notice."
- Initiated
- 2025-05-08
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.
- Recall number
- Z-1980-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213781
- FDA classified
- 2025-06-13
- Recall event number
- 96851