Z-1980-2026: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Recall record Z-1980-2026 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1980-2026
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Apr 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Source facts

Action
Siemens Healthineers notified consignees on about 03/18/2026 via letter. Consignees were notified of the issue, the hazard involved and instructed to immediately discontinue use of and discard the affected kit lots, each facility should review the recall notification with their Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, and review inventory to determine the laboratory's replacement needs and communicate that with Siemens Healthineers. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form provided and retain the letter with laboratory records and forward it to anyone how may have received or used the affected product.
Initiated
2026-03-18
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Recall number
Z-1980-2026
Recall status
Open, Classified
CDRH record ID
219398
FDA classified
2026-04-28
Recall event number
98666

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗