Reason stated in the source
Zimmer recently conducted a review of historic packaging validations completed. Based on this review, it was determined that a subset of product packaged at a single Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a single packaging tray configuration, which
Source facts
- Action
- Zimmer is removing the devices from the field including distributor and hospital inventories. A notification was issued to known users of the product on 07/16/2013.
- Initiated
- 2013-07-16
- Root cause
- Packaging process control
- Terminated
- 2015-04-30
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CT, FL, GA, IL, IN, KS, KY, MA, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VI, and WI, and the countries of Singapore, Canada, Mexico, Arab Emirates, Saudi Arabia, and Hong Kong.
- Recall number
- Z-1981-2013
- Recall status
- Terminated
- CDRH record ID
- 120360
- FDA classified
- 2013-08-17
- Recall event number
- 65833