Skip to content

Public FDA record

Z-1981-2025: UROSKOP Omnia. Model Number: 10094910

Recall record Z-1981-2025 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1981-2025
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jun 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Source facts

Action
Siemens notified consignees on about 06/04/2025 via letter. Consignees were instructed that they can continue to use their system when paying attention for signs of burning smell and/or smoke coming from the device base. In this case, the system must be immediately disconnected from power supply via room emergency on/off switch and the Siemens Healthineers Service Organization must be informed. Siemens Healthineers is preparing an on-site field safety corrective action to replace the frequency inverter by another type that cannot cause the described problem. The field safety corrective action will be implemented in the second quarter of 2025 with update XP008/25/S and will be provided to you free of charge. Consignees were instructed to ensure that all affected personnel are informed as well as any customers if the unit was further distributed and to acknowledge receipt of the recall notification.
Initiated
2025-06-04
Root cause
Device Design
Distribution
Nationwide distribution.
Recall number
Z-1981-2025
Recall status
Open, Classified
CDRH record ID
214357
FDA classified
2025-06-17
Recall event number
97050

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗