Open original FDA record ↗Recall record
- FDA identifier
- Z-1982-2019
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Jul 17, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.
Source facts
- Action
- An "Urgent Medical Device Recall" letter was sent on September 18, 2018 via email to all affected customers. Beckman Coulter ask that if you are using a LIS, please review all previous LIS transmissions which included keywords for accuracy by comparing keywords displayed in Kaluza C Information Plots to those in the LIS system. Please report any erroneous transmissions to Beckman Coulter. The recall letter also ask their customers to visit the Beckman Coulter website to download version Kaluza C 1.1 and install on all affected systems. Installation of Kaluza C version 1.1 will prevent software crashes and result in proper keyword value being transmitted for all samples 1. Navigate to http://www.kaluzasoftware.com and click the "Download" icon. 2. Click "Download Kaluza C1.1" and fill in the requested information. 3. Click the link in the registration email to download Kaluza C version 1.1. 4. Install Kaluza C version 1.1 on systems used for analysis. If you have questions regarding the notification, please contact 1-800-369-0333.
- Initiated
- 2018-09-17
- Root cause
- Software Manufacturing/Software Deployment
- Distribution
- US: CA and NY OUS: United Kingdom, Switzerland and Germany
- Recall number
- Z-1982-2019
- Recall status
- Terminated
- CDRH record ID
- 173179
- FDA record created
- 2019-07-17
- Recall event number
- 82919