Open original FDA record ↗Recall record
- FDA identifier
- Z-1982-2023
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Jun 21, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
Source facts
- Action
- Medline issued a recall notice to its consignees on 05/03/2023 by first calls mail and email The notice explained the problem with the device, risk and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product.2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-23-059 Recall Code: RXXXXXXX 3.Upon completion of the form, please destroy affected product. Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers document and destroy any affected product. You should include your customers quantities on your response form."
- Initiated
- 2023-05-03
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.
- Recall number
- Z-1982-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200304
- FDA classified
- 2023-06-21
- Recall event number
- 92327