Open original FDA record ↗Recall record
- FDA identifier
- Z-1983-2011
- Company in FDA record
- Arjo, Inc. dba ArjoHuntleigh
- FDA classification date
- Apr 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The battery-powered control board in the ceiling lift motor may overheat, resulting in smoke and possibly flames and the lift becoming inoperable.
Source facts
- Action
- ArjoHuntleigh sent an Urgent Field Safety Notice dated March 4, 2011, to all affected customers via UPS Ground on March 21, 2011. The notice informed the accounts of the potential for smoke and flames coming out of the Maxi Sky 1000/V10 bariatric ceiling lifts. The accounts were requested to immediately cease use of the affected lifts, powering them down by activating the emergency stop, and tag the lifts to prevent further use. If there is a clinical need to continue use of the lift before it is retrofitted or replaced, specific instructions were given to reduce the risk of a device malfunction. The accounts were requested to complete the Customer Response Form online, using the enclosed card's instructions. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 57985.
- Initiated
- 2011-03-21
- Root cause
- Device Design
- Terminated
- 2012-02-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-1983-2011
- Recall status
- Terminated
- CDRH record ID
- 98101
- FDA classified
- 2011-04-14
- Recall event number
- 58048