Z-1983-2019: Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664

Recall record Z-1983-2019 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1983-2019
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Jul 17, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.

Source facts

Action
On June 24, 2019, the firm initiated customer notifications in the form of Urgent Medical Device Corrective Action letters. The firm notified users that the following conditions could lead to the workstation becoming unstable and falling over: 1. A monitor weighing more than 19.8 lbs. is attached to the MAJ-1664. 2. The Nurse Control Arm (MAJ-1664) and LCD Monitor Arm (MAJ-1641) are mounted on the WM-NP2/DP2 and are extended in the same direction, which is an unfavorable position in the same direction (e.g. towards the left or right of the workstation). 3. Total weight of the equipment and accessories on the WM-NP/DP2 workstation shelves of less than 66.1lbs 4. When the workstation is stationary on an incline/decline and the brakes are engaged. The communication required customers take the following actions: 1. Review the updated Instructions for Use (Revision Issue 5) regarding the Control Panel Arm and ensure that any monitor placed on the Control Panel Arm does not exceed the maximum loading of 19.8 lbs. 2. If you are currently using a monitor that exceeds the updated maximum loading requirement with the Control Panel Arm, ensure the monitor is removed and replaced with a new monitor that does not exceed 19.8lbs. Please contact your local Olympus Representative for further support. 3. A self-fastening strap should always be used with the Control Panel Arm. Enclosed with the customer notification letter were self-fastening strap(s). Customers were asked to see the updated IFU and apply the supplied self-fastening strap according to the instructions. 4. Inspect your inventory of Instructions for Use (IFU) for the Control Panel Arm and discard any existing IFUs. 5. Olympus recommends that the Control Panel Arm is placed within its storage location and to avoid the four conditions stated within this Field Safety Notice when the WM-NP2/DP2 is transported. 6. Ensure you train your personnel on the updated Instructions for Use for
Initiated
2019-06-10
Root cause
Under Investigation by firm
Terminated
2020-07-10
Distribution
Nationwide domestic distribution.
Recall number
Z-1983-2019
Recall status
Terminated
CDRH record ID
173924
FDA record created
2019-07-17
Recall event number
83171

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗