Skip to content

Public FDA record

Z-1983-2026: BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use

Recall record Z-1983-2026 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1983-2026
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Apr 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.

Source facts

Action
On 04/07/2026, BD sent URGENT: Medical Device Correction letters to customers and distributors. Consignees were instructed that there are no recommendations for retrospective chart review or repeat patient testing. They are to verify incubation atmospheres by incorporating the use of appropriate gas indicators as part of routine laboratory quality control practices. Each BD GasPak EZ Gas Generating Pouch System should be tested periodically for its ability to provide adequate conditions for the growth of target organisms. For CO2 incubation, laboratories may use BD CO2 Indicators (Catalog No. 271055), along with the impacted catalog 260684, to verify sufficient CO2 presence in the environment. If conditions are not confirmed or are suspected to be out of specification, laboratories should repeat incubation using alternative validated methods or established laboratory procedures to ensure organism recovery. Additionally, consignees were instructed to ensure the contents of the notification are read and understood, share this notification with the facility network and forward it to any customers to whom product may have been distributed, and complete and return the Customer Response Form. Distributors were instructed to ensure the notification is read and understood, identify all customers that purchased any affected product, and provide a copy of the attached customer letter to all customers who received the affected lots to advise them of this Product Advisory on BD's behalf. BD is investigating the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD GasPak catalog numbers 260684 and 260685 are currently on backorder due to seal integrity challenges identified in some batches following a recent overhaul of supplier manufacturing equipment. BD anticipates resumption of product shipments by the end of March 2026, with full recovery expected by June 2026.
Initiated
2026-04-07
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the country of Hong Kong.
Recall number
Z-1983-2026
Recall status
Open, Classified
CDRH record ID
219352
FDA classified
2026-04-29
Recall event number
98657

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗