There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
Source facts
Action
Siemens sent a Safety Advisory Notice letter dated May 9, 2014, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that a Siemens service technician would be visiting their site to identify and service affected devices.
Initiated
2014-05-09
Root cause
Component design/selection
Terminated
2014-12-15
Distribution
US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.
Recall number
Z-1985-2014
Recall status
Terminated
CDRH record ID
127494
FDA classified
2014-07-02
Recall event number
68382