Skip to content

Public FDA record

Z-1986-2012: Philips HeartStart MRx M3536A Defibrillator/Monitor with Software R.03.03 The Heartstart MRx is for us

Recall record Z-1986-2012 for Philips Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1986-2012
Company in FDA record
Philips Healthcare Inc.
FDA classification date
Jul 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Defibrillator/monitor cannot analyze an ECG report during acquisition of 12 lead ECG

Source facts

Action
Philips Healthcare issued an Urgent Voluntary Medical Device Correction letter on June 20, 2012, to all affected customers. The Notification informs customers of the issue, identifies details (including models and software versions applicable) of units affected, instructions on actions to take while awaiting the correction, and identifies what action Philips plans to eliminate or remedy the issue A Philips Healthcare representative will contact customers to arrange for the updated R.03.03 software installation in all the affected devices. For questions customers were advised to call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Initiated
2012-06-20
Root cause
Nonconforming Material/Component
Terminated
2013-08-14
Distribution
Worldwide Distribution - USA including FL, IN, MA, M, NY, OH, TN, TX, and VA. Internationally to AUSTRALIA, HONG KONG, and INDIA
Recall number
Z-1986-2012
Recall status
Terminated
CDRH record ID
110296
FDA classified
2012-07-11
Recall event number
62342

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗