Open original FDA record ↗Recall record
- FDA identifier
- Z-1986-2019
- Company in FDA record
- Clariance Inc
- FDA record creation date
- Jul 18, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Potential for the polyaxial screw to become disassembled during implant.
Source facts
- Action
- Clariance, Inc. notified customers on about 07/01/2019 via "MEDICAL DEVICE FIELD SAFETY NOTICE - SN20190603". Customers were informed that when using the 18710024 Screwdriver, the user should ensure that the tip of the screwdriver is properly engaged into the receiving Torx bore of the Erisma-LP polyaxial screw. After proper engagement, the user is able to fully connect the screwdriver on the screw head. Disassembly of the polyaxial screw can occur when the 18710024 Screwdriver is not fully inserted and threaded into the screw head. Additionally when using the 18720000T20 Wrench or 01720001 T20 Snap Fit Tightening Wrench, the user should ensure that the tip of the T20 Wrench or the T20 Snap Fit Tightening Wrench is properly engaged into the receiving Torx bore of the Erisma-LP polyaxial screw. After proper engagement the user is able to adjust the insertion depth of the polyaxial screw. Customers were instructed to read the Field Safety Notice carefully, review the revised Surgical Techniques Guide provided as Attachment A that include in Section 2, Insertion of Pedicle Screws, enhanced instructions to help ensure that the user properly uses the 18710024 Screwdriver, the 18720000T20 Wrench, or the 01720001 T20 Snap Fit Tightening Wrench with the Erisma-LP polyaxial screws, in order to help reduce the possibility of disassembly. They were also instructed to replace the existing Surgical Techniques Guide(s) within your files with the revised Surgical Techniques Guide and to complete and return the response card, Attachment B. Questions can be directed to Clariance via telephone 773-868-7041, or your local sales representative.
- Initiated
- 2019-07-01
- Root cause
- Use error
- Terminated
- 2020-07-09
- Distribution
- Nationwide distribution to CA, IL, IN, NY, OH, TX.
- Recall number
- Z-1986-2019
- Recall status
- Terminated
- CDRH record ID
- 174233
- FDA record created
- 2019-07-18
- Recall event number
- 83218