Skip to content

Public FDA record

Z-1987-2023: Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI

Recall record Z-1987-2023 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1987-2023
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Jun 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

Source facts

Action
Olympus issued Urgent Medical Device Recall Letter dated 5/4/23. Letter states reason for recall, health risk and action to take: 1. Immediately assess any product you have to identify HX-202LR/UR with affected lot number listed in this communication, cease use of product and quarantine any affected product. The images in Attachment 1 depict the area where the lot number is identified. The lot# is on the carton box, pack and handle. 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the recall number: 0424 c. Complete the form as instructed. 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Contact (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Initiated
2023-05-04
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1987-2023
Recall status
Open, Classified
CDRH record ID
200271
FDA classified
2023-06-21
Recall event number
92325

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗