Open original FDA record ↗Recall record
- FDA identifier
- Z-1987-2025
- Company in FDA record
- Preat Corp
- FDA classification date
- Jun 18, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.
Source facts
- Action
- On 04/02/2025 an initial customer notification was distributed by the firm. On 05/05/2025, the firm sent via certified mail an updated "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter to customers informing them that due to manufacturing issues with a dimensional condition that may lead to no engagement between the prosthetic screw and the implant. Customer are instructed to discard or return the affected devices to PREAT Corporation. For questions, contact Quality Assurance at 800-232-7732 or email mgonzales@younginnovations.com
- Initiated
- 2025-04-02
- Root cause
- Process control
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA.
- Recall number
- Z-1987-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213383
- FDA classified
- 2025-06-18
- Recall event number
- 96671