Open original FDA record ↗Recall record
- FDA identifier
- Z-1989-2009
- Company in FDA record
- Breg Inc, An Orthofix Company
- FDA classification date
- Sep 4, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Power supplies short-circuited at the electrical outlet upon initial use of the power supply, causing the circuit breaker for the electrical outlet to trip.
Source facts
- Action
- An Urgent: Medical Device Recall Notification was sent to customers on June 18, 2009. FOR USERS: Customers are advised to immediately open all unused Kodiak cold therapy units and inspect the lot code of the Power Supply included with the Kodiak unit. If the lot code is 3508, 3908, 4308, or 4808, then remove the power supply, cut the cord using scissors and discard. Call Breg Customer Service for a replacement: 1-800-321-0607. Complete and return the response form ASAP. FOR DISTRIBUTORS: Send Breg your distribution information, including contact and quantities. Breg will notify your customers of the recall directly.
- Initiated
- 2009-06-18
- Root cause
- Employee error
- Terminated
- 2011-03-28
- Distribution
- Worldwide Distribution
- Recall number
- Z-1989-2009
- Recall status
- Terminated
- CDRH record ID
- 83978
- FDA classified
- 2009-09-04
- Recall event number
- 52775