Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.
Source facts
Action
Alliance Spine sent an Urgent Medical Device Field Action Notification letter dated June 15, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the response form and fax to 210-314-2524.
Initiated
2015-06-09
Root cause
Device Design
Terminated
2015-10-09
Distribution
US Nationwide Distribution in the states LA, TX, and PR.
Recall number
Z-1989-2015
Recall status
Terminated
CDRH record ID
137878
FDA classified
2015-07-07
Recall event number
71457