Z-1989-2025: Prestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1

Recall record Z-1989-2025 for BALT USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1989-2025
Company in FDA record
BALT USA, LLC
FDA classification date
Jun 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

Source facts

Action
On 05/08/2025, the firm sent via priority mail and an, "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that defect impacting specific lots of the Optima and Prestige Coil Systems, in which the distal 3 cm radiopaque (RO) marker may not be visible under fluoroscopy. Customers are instructed to: As a voluntary alternative to returning the affected products, devices from the affected lots can be inspected by trained hospital personnel for the presence of the RO marker using available fluoroscopy imaging equipment prior to a clinical procedure. It is important to conduct this inspection as soon as possible. The image is achieved by taking the fluoroscopic image of the coil system within the unopened box. 1) If the image contains the RO marker, as exemplified from Figure 1, i. Complete the Notice of Receipt (refer to the annex on page 5) indicating the verified results. ii. Mark the inspected boxes with the date and time of the fluoroscopy test and the name of the person performing the test. iii. Return the product back to your inventory with no further action required. iv. Send the completed Notice of Receipt to Balt USA at QA@baltgroup.com. 2) If the image does not contain the RO marker from Figure 2 or if the RO marker is not visually discernable, contact Balt USA at QA@baltgroup.com to process the return of the product. If customers desire to return the products: Distribution agents or Hospital Staff: -Identify and locate the affected Coil System and cease all use of the affected product(s). -For returning collect and put in quarantine and then return them to Balt USA through the usual RMA procedure by contacting our Customer Service. -Contact Balt USA at QA@baltgroup.com or phone +949-788-1443.
Initiated
2025-05-08
Root cause
Vendor change control
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IL, IN, KS, KY, MA, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA and WI. The countries of Argentina, Armenia, Australia, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Czechia, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Italy, Japan, Jordan, Kazakhstan, Korea, Kyrgyzstan, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Peru, Poland, Portugal, Romania, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Uruguay, and Vietnam.
Recall number
Z-1989-2025
Recall status
Open, Classified
CDRH record ID
213803
FDA classified
2025-06-18
Recall event number
96787

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗