Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).
Source facts
Action
Olympus Scientific Solutions planned action:
1. Notify product managers as to potential issues
2. Distribute new table and notice to instrument owners 30 JUL 15
3. Post new table on website for access 15 JUN 15
Initiated
2015-06-09
Root cause
Component change control
Distribution
Nationwide Distribution.
Recall number
Z-1991-2015
Recall status
Open, Classified
CDRH record ID
138532
FDA classified
2015-07-31
Recall event number
71649