Open original FDA record ↗Recall record
- FDA identifier
- Z-1991-2021
- Company in FDA record
- Bard Access Systems Inc.
- FDA record creation date
- Jun 30, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Catheter kits contained the incorrect micro-introducers.
Source facts
- Action
- On 05/19/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via FedEx second day air to customer informing them that they have confirm through customer complaints that catheter kits contains the incorrect micro-introducers. Customer are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers and destroy all product subject to the recall following their institutions process for destruction. The Recalling Firm will replace discarded product. 2. Share this recall notification with all users within or outside their facility network to which they may have further distributed the affected products to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not they have any of the impacted material so that Recalling Firm may acknowledge the receipt of this notification per FDA requirements and subsequently process your product replacement. For any questions or further assistance - contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
- Initiated
- 2021-05-19
- Root cause
- Process control
- Terminated
- 2023-01-24
- Distribution
- U.S.: CT, GA, MA, MI, MT, NJ, NV, NY, VA, and WI. O.U.S.: None
- Recall number
- Z-1991-2021
- Recall status
- Terminated
- CDRH record ID
- 187775
- FDA record created
- 2021-06-30
- Recall event number
- 88048