Product devices used for testing are not identical to what was manufactured. Under certain circumstances snap feature could allow poly liner to move on baseplate reducing tightness of fit.
Source facts
Action
Firm notified consignees by an Urgent: Medical Device Recall letter, dated July 20, 2009. The letter stated the reason for the recall, asked customers to quarantine, and return the affected product. The enclosed response form needed to be completed and returned as soon as possible. Questions were to be directed to David Harris, Manager of Customer Service at 512-834-6330.
Initiated
2009-07-20
Root cause
Release of Material/Component prior to receiving test results
Terminated
2010-01-07
Distribution
Nationwide Distribution -- Including states of TX, FL, and GA.
Recall number
Z-1992-2009
Recall status
Terminated
CDRH record ID
83676
FDA classified
2009-09-04
Recall event number
52701