Open original FDA record ↗Recall record
- FDA identifier
- Z-1992-2020
- Company in FDA record
- Covidien Llc
- FDA record creation date
- May 14, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Source facts
- Action
- Cardinal Health issued letter Urgent Medical Device Recall EVENT-2020-02846 on 4/30/20 via overnight mail. Letter states reason for recall, health risk, and action to take: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers 2. SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Exhibit A. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: Hospital800-964-5227 Federal Government800-444-1166 Distributor800-635-6021 All other Customers888-444-5440 Questions contact: Cardinal Health 800-292-9332
- Initiated
- 2020-04-30
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-12-14
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1992-2020
- Recall status
- Terminated
- CDRH record ID
- 181419
- FDA record created
- 2020-05-14
- Recall event number
- 85629