Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than 35 percent. The dosage displayed on the generator console and table display (monitor) is higher than the actual delivered dose.
Source facts
Action
The product can be brought into compliance via a software update. The firm plan to have the updated software installed, working with customer schedules, over the course of the next 12 months.
Initiated
2021-03-11
Root cause
Radiation Control for Health and Safety Act
Terminated
2024-09-18
Distribution
US Nationwide Distribution and in the country of Canada
Recall number
Z-1992-2021
Recall status
Terminated
CDRH record ID
188065
FDA record created
2021-07-02
Recall event number
88195