Z-1992-2023: Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW8

Recall record Z-1992-2023 for Draeger Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1992-2023
Company in FDA record
Draeger Medical Systems, Inc.
FDA classification date
Jun 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.

Source facts

Action
An Urgent - Medical Device Correction dated May 2023 was sent to customers. Actions to be taken: Please ensure that all users of the above-mentioned products and other persons within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information. Users can mitigate this risk by performing the visual check, test procedure and take appropriate actions based on the Attachment provided in this notice. Complete and return the enclosed Customer Acknowledgement and Response Form . We are continuing to evaluate this issue and will advise the affected customers when we have an alternative solution. Visual Check for Correct Mattress Placement: To ensure proper operation of the scale, please check to confirm there is a clearance between the mattress cover and the corner blocks and side panels of the Resuscitaire. 1. Ensure mattress with cover is installed in the Resuscitaire bassinet. 2. Observe edges of mattress for clearance with corner blocks and side panels, see Figure 2. 3. If there is interference, then the scale may not provide accurate weight measurements. It is recommended to use an independent scale for therapeutic decisions and treatment Test Procedure If it is unclear from the visual check if the mattress cover interferes with the corner blocks and/or side panels, impacting scale function, please perform the following test: a. Place a known calibrated weight on the center mattress (with mattress cover installed). A 5kg weight is recommended in the instructions for use. b. Execute weighing procedure and record weight. c. Repeat weighing procedure 5-10 times. d. Compare measurements. i. If measurements are within 5 grams of calibrated 5kg weight, mattress assembly is acceptable. ii. If measurements deviate more than +/- 5 grams from the calibrated 5kg weight, it is recommended to use an alternate weighing scale Contact: We regret any inconvenience this may cause.
Initiated
2023-05-16
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, VT, WA, and WI.
Recall number
Z-1992-2023
Recall status
Open, Classified
CDRH record ID
200459
FDA classified
2023-06-22
Recall event number
92353

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗