Open original FDA record ↗Recall record
- FDA identifier
- Z-1992-2025
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Jun 18, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
Source facts
- Action
- Fresenius Medical Care notified its single consignee on 04/29/2025 via telephone call. The consignee was made aware of the issue and affected units and instructed to examine their inventory for those affected units. If they had any affected units, they were instructed to remove them from their inventory immediately and provide contact information to arrange for their return. On 05/14/2025, Fresenius Medical Care expanded the recall to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. On 06/05/2025, Fresenius Medical Care expanded the recall again to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. A recall notification letter was sent to the consignee on about 06/23/2025. The letter provided the same information as was verbally communicated in notifications.
- Initiated
- 2025-04-29
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the state of Massachusetts.
- Recall number
- Z-1992-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213736
- FDA classified
- 2025-06-18
- Recall event number
- 96827