Skip to content

Public FDA record

Z-1993-2023: HeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore

Recall record Z-1993-2023 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1993-2023
Company in FDA record
Heartware, Inc.
FDA classification date
Jun 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal.

Source facts

Action
Medtronic Mechanical Circulatory Support issued an URGENT MEDICAL DEVICE COMMUNICATION notice to it consignees on 05/16/2023 by email. The notice explained that logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal. Medtronic is working on an update to the Autologs web portal to resolve the issue. Until the update is complete, logfiles can be sent to dl.mcsclinicalengineering@medtronic.com for report generation. Please note in the email that you have received a logfile upload error and are requesting a logfile report. On about 09/20/2023, a formal Urgent Medical Device Communication notification was sent to Healthcare Professionals (HCPs) informing them of the resolution of the autologs web portal issue previously communicated. Consignees are asked to confirm receipt of the letter and that they have acknowledged the actions to be taken.
Initiated
2023-05-16
Root cause
Software change control
Distribution
Worldwide - US Nationwide distribution in the states of TX, MO, VA, and the countries of Bahrain, UK.
Recall number
Z-1993-2023
Recall status
Open, Classified
CDRH record ID
200448
FDA classified
2023-06-22
Recall event number
92350

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗