Open original FDA record ↗Recall record
- FDA identifier
- Z-1993-2023
- Company in FDA record
- Heartware, Inc.
- FDA classification date
- Jun 22, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal.
Source facts
- Action
- Medtronic Mechanical Circulatory Support issued an URGENT MEDICAL DEVICE COMMUNICATION notice to it consignees on 05/16/2023 by email. The notice explained that logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal. Medtronic is working on an update to the Autologs web portal to resolve the issue. Until the update is complete, logfiles can be sent to dl.mcsclinicalengineering@medtronic.com for report generation. Please note in the email that you have received a logfile upload error and are requesting a logfile report. On about 09/20/2023, a formal Urgent Medical Device Communication notification was sent to Healthcare Professionals (HCPs) informing them of the resolution of the autologs web portal issue previously communicated. Consignees are asked to confirm receipt of the letter and that they have acknowledged the actions to be taken.
- Initiated
- 2023-05-16
- Root cause
- Software change control
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, MO, VA, and the countries of Bahrain, UK.
- Recall number
- Z-1993-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200448
- FDA classified
- 2023-06-22
- Recall event number
- 92350