Patient Side Cart battery boxes may overheat in rare charging conditions, causing battery bulging.
Source facts
Action
No notification was sent. Issue is being addressed during routine maintenance or through complaint process.
Initiated
2012-04-27
Root cause
Device Design
Terminated
2014-07-07
Distribution
Worldwide Distribution.
Recall number
Z-1994-2014
Recall status
Terminated
CDRH record ID
128285
FDA classified
2014-07-04
Recall event number
68593
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.