Catheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire.

Source facts

Action
On 05/19/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via FedEx 2-day to customers informing them that the Recalling Firm has confirmed through customer complaint that a number of catheter convenience kits labeled as having 70cm guidewires may actually contain 50cm guidewires. The Recalling Firm is providing customers with a quick way of identifying if their kit contains the correct or incorrect size guidewire by determining the number of loops of the wire. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Determine if the product has the correct product based on the number of loops of the guidewire while the product is still in the packaging material. Product may be discarded when clinically necessary and replacement requested. 2. Share this recall notification with all users within or outside their facility network to which they may have further distributed the affected products to ensure these additional users are also aware of this advisory notice. 3. Complete the attached Customer Response Form and return to the Recalling Firm contact noted on the form whether or not they have any of the impacted material so that Recalling Firm may acknowledge receipt of this notification per FDA requirements and subsequently process the product replacement, if necessary. For further assistance or for questions - contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
Initiated
2021-05-19
Root cause
Process control
Terminated
2023-03-31
Distribution
U.S. Nationwide distribution in the state of RI.
Recall number
Z-1994-2021
Recall status
Terminated
CDRH record ID
187797
FDA record created
2021-07-01
Recall event number
88057

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗