Open original FDA record ↗Recall record
- FDA identifier
- Z-1995-2013
- Company in FDA record
- Blue Belt Technologies MN
- FDA classification date
- Aug 21, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Blue Belt Technologies is recalling sterile Monitor Drapes used with their Navio PFS system during orthopedic knee surgery. The drapes are used to cover the Navio PFS system computer monitor and arm assembly and provide a sterile barrier in the surgical suite. The defect is evident when the computer monitor is being draped. The front section of the drape separates from the rest of the drape at
Source facts
- Action
- Blue Belt Technologies contacted consignees by telephone using a script on July 19, 2013. The consignees were informed of the problem and the product involved in the recall. Consignees were advised to immediate stop using the product and to locate and quarantine all unused product. Additionally, an "Urgent Medical Device Recall" letter dated July 24, 2013 was issued on July 24, 2013 and August 12, 2013. The letter includes the description of the Issue, description of safety concerns, affected product, and action to be taken by user. For questions contact Blue Belt Technologies, Lindsey Bring (+1 763.452.4922) or Robin Gonzalez (+1 763.452.4910) and get a Return Authorization number. Lindsey or Robin will arrange a pickup of the product. BBT will send replacement product at no charge.
- Initiated
- 2013-07-19
- Root cause
- Package design/selection
- Terminated
- 2013-11-14
- Distribution
- Worldwide Distribution and US Nationwide in the states of: CA and PA. United Kingdom.
- Recall number
- Z-1995-2013
- Recall status
- Terminated
- CDRH record ID
- 120387
- FDA classified
- 2013-08-21
- Recall event number
- 65834