- FDA identifier
- Z-1995-2014
- Company in FDA record
- Cenorin, LLC
- FDA classification date
- Jul 6, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
Source facts
- Action
- Cenorin sent a Notification letter dated December 18, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated: Cenorin has developed an upgrade to mitigate this issue as well as provide additional process control and system monitoring benefits. Cenorin included a Return Response form with the recall letter to be completed and return. For questions contact HLD Systems Sales Representative for questions.
- Initiated
- 2013-01-24
- Root cause
- Device Design
- Terminated
- 2015-02-10
- Distribution
- Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
- Recall number
- Z-1995-2014
- Recall status
- Terminated
- CDRH record ID
- 128261
- FDA classified
- 2014-07-06
- Recall event number
- 68580