Z-1995-2020: Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with Stea

Recall record Z-1995-2020 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1995-2020
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
May 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).

Source facts

Action
Urgent Medical Device Correction Letters were sent to U.S. customers via FedEx Overnight with signature required at delivery beginning on August 9, 2019. European consignees were first notified via registered mail and/or personal visits by Medtronic representatives starting August 27, 2019. The one Canadian consignee was first notified via registered mail starting August 19, 2019. Customers were informed of which devices, when used in combination, can result in entry point and lead placement inaccuracies during Deep Brain Stimulation (DBS) lead implantation procedures. Customers were instructed to: 1. determine which registration method, auto-registration (fiducial-less) or manual registration, is appropriate considering the information in the Urgent Medical Device Correction Letter, 2. assess navigational accuracy on several known anatomical landmarks before using the registration for navigation, 3. use the StealthMerge functionality in the software to compare the actual location of the cannula or lead to the surgical plan, 4. use the planning functionality in the software to compare the cannula to planned trajectory, 5) sign and date the Customer Confirmation Form, and return the form to Medtronic via email at RS.NavFCA@Medtronic.com or via fax at 651-367-7075 within 30 days of receipt. Neurosurgeon users were informed that a Medtronic representative would contact them regarding training available to provide an overview of this issue. Users were requested to register for and complete the training. Customers were instructed to report adverse events or quality problems experienced with the product to FDA and Medtronic as follows: E-mail Medtronic at rs.navtechsupport@medtronic.com or call 1-888-826-5603; Online at the FDA website2 (form available to fax or mail) or call FDA at 1-800-FDA-1088. Customers with questions were instructed to contact Medtronic at 1- 888-826-5603.
Initiated
2019-08-09
Root cause
Software design
Terminated
2024-04-02
Distribution
Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom.
Recall number
Z-1995-2020
Recall status
Terminated
CDRH record ID
180357
FDA classified
2020-05-18
Recall event number
85243

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗