Skip to content

Public FDA record

Z-1995-2021: MaxPlusTM bi fuse extension set with 2 clear needleless connectors - Product Usage: The needleless conne

Recall record Z-1995-2021 for CAREFUSION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1995-2021
Company in FDA record
CAREFUSION
FDA record creation date
Jul 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A portion of a validation lot was inadvertently released to the US Market. The product did not meet the validation testing criteria regarding leakage.

Source facts

Action
On 06/09/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via email and FedEx to customers informing them that a validation lot was inadvertently released to the US market. This product did not meet the validation testing criteria regarding leakage. Potential health consequences associated with the use of the non-validated product is under-infusion. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Destroy all product subject to the recall following their institutions process for destruction. The Recalling Firm will replace discarded product. 2. Share this recall notification with all users within or outside their facility network to which you they have further distributed the affected products to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the Recalling Firm's contact noted on the form (BD Post Market Quality Email: BDRC9@bd.com, Or Fax No.: 312-949-0236) whether or not they have any of the impacted material so that the Recalling Firm may acknowledge their receipt of this notification per FDA requirements and subsequently process their product replacement. For any questions or assistance contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
Initiated
2021-06-09
Root cause
Under Investigation by firm
Terminated
2024-07-08
Distribution
U.S.: TN O.U.S.: None
Recall number
Z-1995-2021
Recall status
Terminated
CDRH record ID
187866
FDA record created
2021-07-01
Recall event number
88104

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗