Open original FDA record ↗Recall record
- FDA identifier
- Z-1996-2010
- Company in FDA record
- Abbott Point of Care Inc.
- FDA classification date
- Jul 13, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.
Source facts
- Action
- Urgent recall notifications were sent on May 13 2010. Customers were instructed to discontinue use and to return the identified lots. Additional questions are directed to Abbott Point of Care Technical Support at 800-3668020, Option 1 or your Abbott Point of Care representative.
- Initiated
- 2010-05-05
- Root cause
- Other
- Terminated
- 2011-09-22
- Distribution
- Worldwide distribution: Argentina, Hong Kong, China, Finland, Greece, India, Portugal, Russian Federation, Saudi Arabia, United Arab Emirates, Italy, Spain, Australia, Germany, UK, Sweden, The Netherlands, France, Canada.
- Recall number
- Z-1996-2010
- Recall status
- Terminated
- CDRH record ID
- 92194
- FDA classified
- 2010-07-13
- Recall event number
- 55895