Skip to content

Public FDA record

Z-1996-2021: Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra P

Recall record Z-1996-2021 for Flowonix Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1996-2021
Company in FDA record
Flowonix Medical Inc
FDA record creation date
Jul 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection

Source facts

Action
Flowonix Customer letters) have been mailed via USPS 2 day priority mail with electronic tracking on 5/26-5/27. Letter states reason for recall, health risk and Actions to be taken by the Customer: 1) Discontinue the use and distribution of the Physician Order Form (PL-15400-00, Nov. 2018). This form is no longer available for use and shall no longer be used. 2) Complete and return the attached response form to Flowonix. Should you have any questions, please contact your Flowonix Representative or our Technical Solutions Department at (855) 356-9665
Initiated
2021-05-26
Root cause
Error in labeling
Terminated
2024-02-20
Distribution
US Nationwide distribution.
Recall number
Z-1996-2021
Recall status
Terminated
CDRH record ID
187755
FDA record created
2021-07-01
Recall event number
88044

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗