Z-1996-2025: DxI 9000 Access Immunoassay Analyzer C11137

Recall record Z-1996-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1996-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jun 20, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.

Source facts

Action
The firm began notifying customers the week of May 7, 2025 via Urgent Medical Device Recall letters. Only analyzers running software version 1.20 are affected. Customers were provided with instructions on what to do if they observe a calibration that remains in the "In Progress" or "Presented" state. Beckman Coulter will release a service modification to change a configuration setting to prevent this issue. Your service representative will schedule the service modification installation when it is available. In the meantime, it is recommended that customers share the content of this letter with your laboratory and/or medical director to determine if a review of previous patient test results is necessary. Beckman Coulter recommends posting the recall letter on or near the affected systems until corrections have been made.
Initiated
2025-05-07
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Libya, New Zealand, Oman, Poland, Portugal, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-1996-2025
Recall status
Open, Classified
CDRH record ID
213852
FDA classified
2025-06-20
Recall event number
96876

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗