Skip to content

Public FDA record

Z-1996-2026: iLet Bionic Pancreas, REF: BB1001

Recall record Z-1996-2026 for Beta Bionics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1996-2026
Company in FDA record
Beta Bionics, Inc.
FDA classification date
Apr 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.

Source facts

Action
On 3/12/2026, correction notices were emailed to customers who were asked to do the following: Update device software to host software 1.4.4 and up. Update your software 1. Download or update your iLet app on your phone. Read the iLet App User Guide: https://www.betabionics.com/wp-content/uploads/LA000056_F-iLet-Mobile-App-User-Guide.pdf 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. -Distributors are asked to provide patient contact information so the firm can provide the correction notice to these patients. - Complete the acknowledgment survey https://sprw.io/stt-6HnQU If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.
Initiated
2026-03-12
Root cause
Software change control
Distribution
US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Recall number
Z-1996-2026
Recall status
Open, Classified
CDRH record ID
218863
FDA classified
2026-04-29
Recall event number
98475

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗