Toshiba America Medical Systems is initiating a field correction on the PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems due to an incorrect sweep direction setting. Images acquired using the endo-cavitary transducer PVT-681MV and cross-sectional images created from these volume images are reversed
horizontally on the Aplio XG (SSA-790A) and the Xario XG (
Source facts
Action
Toshiba America Medical Systems (TAMS) will issue notification to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification. matter. Questions regarding this letter are directed to the firm at (800) 421-1968 or to your local Toshiba representative.
Initiated
2010-01-05
Root cause
Other
Terminated
2012-02-29
Distribution
Nationwide
Recall number
Z-1997-2010
Recall status
Terminated
CDRH record ID
91088
FDA classified
2010-07-13
Recall event number
55544