Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1).
When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.
Source facts
Action
On November 30, 2018, Philips initiated a software update in the field. Among other performance improvements , this software update included the
correction for poor image quality under very specific situations.
Initiated
2018-11-30
Root cause
Device Design
Terminated
2024-03-27
Distribution
Nationwide
Recall number
Z-1997-2019
Recall status
Terminated
CDRH record ID
174975
FDA record created
2019-07-30
Recall event number
83366