Z-1997-2021: Tempus Pro - Patient physiological monitor, arrhythmia detector and alarm. Part No. - 00-1004-R, 00-1007

Recall record Z-1997-2021 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1997-2021
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA record creation date
Jul 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Tempus Pro (Trizeps 7 only) when used in combination with a specified laryngoscope device can cause two error types when subsequently unplugged.

Source facts

Action
On June 2, 2021, Philips issued an Urgent Medical Device Correction notice to customers via letter and email notifying them an intermittent software error has been detected as a part of an internal production control activity. The error can occur immediately after unplugging the Video Laryngoscope from the USB port. If the customer identifies affected products in use or in inventory: Clinical User Actions: - Users of Tempus Pro who are not using a Video Laryngoscope do not need to take any further actions. - When a Video Laryngoscope is required to be used, the following actions are recommended to reduce the probability of a software reboot during the intervention; - An alternative Video Laryngoscope device may be used to perform any critical interventions such as intubation. - If you are using the Tempus Pro and need to use the Video Laryngoscope on a patient, we recommend completion of any critical interventions before unplugging the Video Laryngoscope from the USB port. - Should the Video Laryngoscope image become frozen on the screen, re-start (turn off and turn back on) the Tempus Pro device. Customer Actions: - Notify all users of the potential risk of loss of monitoring when unplugging the Video Laryngoscope from the Tempus Pro, requiring the device to be restarted. - This issue is limited to specific serial numbers. Please verify that your device is associated with the product included in the scope of this notice utilising attachment 1. - Return your completed Customer Reply form to the email address below, to allow RDT to provide the software update: - Email completed and signed form to: RDT.Recall.Response@Philips.com
Initiated
2021-05-25
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, TX, PA, MN, UT, MT, CA, WI, VA, NY, TN, KS, MD and the countries of Australia, Austria, Czech Republic, France, Germany, Hong Kong, Israel, Italy, Netherlands, Thailand, Abu Dhabi, UK.
Recall number
Z-1997-2021
Recall status
Open, Classified
CDRH record ID
187783
FDA record created
2021-07-01
Recall event number
88051

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗