Z-1997-2025: Bigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Relea

Recall record Z-1997-2025 for BIGFOOT BIOMEDICAL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1997-2025
Company in FDA record
BIGFOOT BIOMEDICAL
FDA classification date
Jun 20, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision.

Source facts

Action
On 05/21/2025, the firm sent via FedEx and USPS Priority Mail, "URGENT MEDICAL DEVICE RECALL" Letters to Customers and Prescribers informing them that the Bigfoot Unity Pen Caps, both Long-Acting and Rapid-Acting Pen Cap firmware version 1.3.0 may provide a screen with "Device Error - Contact Bigfoot Customer Care" and render the pen caps unresponsive and unable to function. Customer are instructed to: 1. If customers have the affected product identified above, we are asking you to return the product. We have sent you a return kit, which you should receive shortly. If the return kit does not arrive in the next 7 days, please contact Bigfoot Biomedical Customer Care at 1-551-244-3668 or 1-888-523-3662 (toll-free number). 2. Ensure you have and continue to use a backup diabetes management plan and diabetes supplies if your Bigfoot Unity Pen Cap stops working. 3. As soon as possible, complete and return the Acknowledgement Postcard to acknowledge the understanding of the actions to be taken. o If customer's device experiences Device Error or other product issues in the future or have any questions, contact Bigfoot Biomedical Customer Care at 1-551-244-3668 or 1-888-523-3662 (tollfree number).
Initiated
2025-05-21
Root cause
Software design
Distribution
U.S. Nationwide distribution in the states of AL, FL, GA, IA, IL, NY, PA, TX, and WA.
Recall number
Z-1997-2025
Recall status
Open, Classified
CDRH record ID
214046
FDA classified
2025-06-20
Recall event number
96925

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗