Open original FDA record ↗Recall record
- FDA identifier
- Z-1999-2011
- Company in FDA record
- Hospira Inc.
- FDA classification date
- Apr 15, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The software enables programming of the GemStar device for a Bolus Lower Lockout Limit outside of the hard limits entered into the GemStar SP Infusion Suite Database PC application.
Source facts
- Action
- The firm, Hospira, sent "URGENT DEVICE FIELD CORRECTION" letters dated March 23, 2011 to their customers on the same date. The letter described the product, problem and preventative action to be taken by the customers until the software can be upgraded. The customers were requested to contact Hospira Global Product Safety and Complaints at 1-800-441-4100 to report adverse events, and Hospira Advanced Knowledge Center at 1-800-241-4002 for technical assistance. The accounts were also requested to complete the attached Reply Form and return it to Hospira via fax at 1-262-577-6921. For additional information and technical assistance, contact Hospira Advanced knowledge Center at 1-800-241-4002, option 4 (available 24 hours a day/7 days per week).
- Initiated
- 2011-03-23
- Root cause
- Software design
- Terminated
- 2016-08-29
- Distribution
- Worldwide distribution: USA (nationwide) and countries including: Canada, Chile, Colombia and Singapore
- Recall number
- Z-1999-2011
- Recall status
- Terminated
- CDRH record ID
- 98667
- FDA classified
- 2011-04-15
- Recall event number
- 58202